A 95 Peptidstørrelsesfordeling – Validering ny kolonne

Bidragsytere:
  • Tor-Arne Krakeli

Utgiver/serie

Utgiver

Nofima AS

Serie

Nofima rapportserie
ISSN 1890-579X

Om resultatet

Rapport
Publiseringsår: 2021
Hefte: 36/2021
Antall sider: 19
ISBN: 978-82-8296-699-3
Open Access

Beskrivelse Beskrivelse

Tittel

A 95 Peptidstørrelsesfordeling – Validering ny kolonne

Tittel

A 95 Molecular Weight Distribution – Validation of the new column

Sammendrag

In the analysis ‘A 95 Molweight Distribution of Peptides’ BioLab use the Size Exclusion Chromatography (SEC) column Superdex Peptide. After a request for a new column of the same type it was informed that the column was no longer produced. The supplier recommended a replacement column, Superdex Increase. After a validation where the two columns was compared, it was concluded that BioLab could use Superdex Increase as a replacement for Superdex Peptide. This lead to a slight increase in the uncertainty for the methode, and a restriction in the measuring area. From ‘

Bidragsytere

Tor-Arne Krakeli

  • Tilknyttet:
    Forfatter
    ved BioLab ved NOFIMA
1 - 1 av 1