Background:
Controversy exists regarding surveillance of arteriovenous fistulas for hemodialysis to increase patency. A significant reduction in volume flow rate (VFR) should lead to diagnostic evaluation and eventually intervention. Several methods are available for VFR measurements, but all of them are associated with low reproducibility. VFR trend analysis is suggested as an improved solution. It is therefore a need to find user-friendly, cost and time-effective modalities. We present a novel Doppler ultrasound device (earlybird) which could bridge this gap. It includes an easy-to-use and light-weight single element transducer.
Methods:
In an experimental and clinical setting, we compared earlybird to duplex ultrasound to assess VFR. In a closed circuit of blood-mimicking fluid, 36 paired calculations of calibrated, duplex ultrasound and earlybird VFR was measured. In addition, 23 paired recordings of duplex ultrasound and earlybird VFR was measured in 16 patients with underarm arteriovenous fistulas. Pearson correlation, intraclass correlation coefficient, root-mean-square and Bland-Altman plots were analyzed.
Results:
Strong correlation (r = 0.991, p