Sammendrag
Cardiac power, the product of aortic flow and blood pressure, appears to be a fundamental cardiovascular parameter. The simplified version named cardiac power output (CPO), calculated as the product of cardiac output (CO) in L/min and mean arterial pressure (MAP) in mmHg divided by 451, has shown great ability to predict outcome in a broad spectrum of cardiac disease. Beat-by-beat evaluation of cardiac power (PWR) therefore appears to be a possibly valuable addition when monitoring circulatory unstable patients, providing parameters of overall cardiovascular function. We have developed a minimally invasive system for cardiac power measurement, and aimed in this study to compare this system to an invasive method (ttPWR). Seven male anesthetized farm pigs were included. A laptop with in-house software gathered audio from Doppler signals of aortic flow and blood pressure from the patient monitor to continuously calculate and display a minimally invasive cardiac power trace (uPWR). The time integral per cardiac cycle (uPWR-integral) represents cardiac work, and was compared to the invasive counterpart (ttPWR-integral). Signals were obtained at baseline, during mechanically manipulated preload and afterload, before and after induced global ischemic left ventricular dysfunction. We found that the uPWR-integral overestimated compared to the ttPWR-integral by about 10% (P
Vis fullstendig beskrivelse