Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film

Bidragsytere:
  • Bjørn Gjerdrum
  • Kjell Gunnar Gundersen
  • Per Olof Lundmark og
  • Bente Monica Aakre

Tidsskrift

Clinical Ophthalmology
ISSN 1177-5467
e-ISSN 1177-5483
NVI-nivå 1

Om resultatet

Vitenskapelig artikkel
Publiseringsår: 2020
Trykket: 2020
Hefte: 14
Sider: 3991 - 4003
Open Access

Importkilder

Scopus-ID: 2-s2.0-85098252672

Beskrivelse Beskrivelse

Tittel

Repeatability of OCT-Based versus Scheimpflug- and Reflection-Based Keratometry in Patients with Hyperosmolar and Normal Tear Film

Sammendrag

Purpose: To compare the repeatability of keratometry between different instruments in patients with hyperosmolar tear film and a control group. Patients and Methods: Subjects with tear-film osmolarity of 316 mOsm/L or more in either eye or 308 m/Osm/L or lower in both eyes were assigned to the hyperosmolar and the control group, respectively. The test eye was the eye with higher osmolarity in the hyperosmolar group and randomly chosen in the control group. The repeatability of keratometry was compared between a reflectometry device (Haag-Streit Lenstar 900), a Scheimpflug device (Oculus Pentacam HR) and two optical coherence tomography (OCT) devices (Tomey Casia SS-1000 and Heidelberg Anterion), based on two measurements from each device. Results: The study included 94 subjects (31 hyperosmolar and 63 controls). Both OCT devices had higher mean differences of average simulated keratometry (SimK) vs the Lenstar in both groups, though all differences in means were

Bidragsytere

Bjørn Gjerdrum

  • Tilknyttet:
    Forfatter
    ved Institutt for optometri, radiografi og lysdesign ved Universitetet i Sørøst-Norge
  • Tilknyttet:
    Forfatter
    ved IFocus Øyeklinikk AS

Kjell Gunnar Gundersen

  • Tilknyttet:
    Forfatter
    ved IFocus Øyeklinikk AS

Per Olof Lundmark

  • Tilknyttet:
    Forfatter
    ved Institutt for optometri, radiografi og lysdesign ved Universitetet i Sørøst-Norge

Bente Monica Aakre

  • Tilknyttet:
    Forfatter
    ved Institutt for optometri, radiografi og lysdesign ved Universitetet i Sørøst-Norge
1 - 4 av 4