Continuous fetal cerebral blood flow monitoring during labor: A feasibility study

Bidragsytere:
  • Kristin Melheim Strand
  • Hans Garman Torp
  • Anne Engtrø Husby
  • Kjell Åsmund Blix Salvesen og
  • Siri Ann Nyrnes

Tidsskrift

Early Human Development
ISSN 0378-3782
e-ISSN 1872-6232
NVI-nivå 1

Om resultatet

Vitenskapelig artikkel
Publiseringsår: 2023
Publisert online: 2023
Trykket: 2023
Volum: 182

Importkilder

Scopus-ID: 2-s2.0-85160677448

Beskrivelse Beskrivelse

Tittel

Continuous fetal cerebral blood flow monitoring during labor: A feasibility study

Sammendrag

Background Current methods for fetal surveillance during labor have significant limitations. Since continuous fetal cerebral blood flow velocity (CBFV) monitoring during labor may add valuable information about fetal well-being, we developed a new ultrasound system called VisiBeam. VisiBeam consists of a flat probe (diameter 11 mm) with a cylindric plane wave beam, a vacuum attachment (diameter 40 mm), a scanner, and a display. Aims To assess the feasibility of VisiBeam for continuous fetal CBFV monitoring during labor, and to study changes in CBFV during uterine contractions. Study design Descriptive observational study. Subjects Twenty-five healthy women in labor with a singleton fetus in cephalic presentation at term. A transducer was placed over a fontanelle and attached to the fetal head with vacuum suction. Outcome measures Achievement of continuous good quality fetal CBFV measures, such as peak systolic velocity, time averaged maximum velocity and end diastolic velocity. Trend plots of velocity measures display changes in CBFV between and during uterine contractions. Results Good quality recordings during and between contractions were achieved in 16/25 fetuses. In twelve fetuses, CBFV measures were stable during uterine contractions. Four fetuses showed patterns of reduced CBFV velocity measures during contractions. Conclusions Continuous fetal CBFV monitoring by VisiBeam was feasible in 64 % of the subjects during labor. The system displayed variations of fetal CBFV not available by today's monitoring techniques and motivates for further studies. However, improvement of the probe attachment is required to ensure good quality signal in a higher proportion of fetuses during labor.

Bidragsytere

Kristin Melheim Strand

  • Tilknyttet:
    Forfatter
    ved Institutt for sirkulasjon og bildediagnostikk ved Norges teknisk-naturvitenskapelige universitet
  • Tilknyttet:
    Forfatter
    ved Kvinneklinikken ved St. Olavs Hospital HF

Hans Torp

Bidragsyterens navn vises på dette resultatet som Hans Garman Torp
  • Tilknyttet:
    Forfatter
    ved Institutt for sirkulasjon og bildediagnostikk ved Norges teknisk-naturvitenskapelige universitet

Anne Engtrø Husby

  • Tilknyttet:
    Forfatter
    ved Kvinneklinikken ved St. Olavs Hospital HF
  • Tilknyttet:
    Forfatter
    ved Institutt for klinisk og molekylær medisin ved Norges teknisk-naturvitenskapelige universitet

Kjell Åsmund Blix Salvesen

  • Tilknyttet:
    Forfatter
    ved Kvinneklinikken ved St. Olavs Hospital HF
  • Tilknyttet:
    Forfatter
    ved Institutt for klinisk og molekylær medisin ved Norges teknisk-naturvitenskapelige universitet

Siri Ann Nyrnes

  • Tilknyttet:
    Forfatter
    ved Barne- og ungdomsklinikken ved St. Olavs Hospital HF
  • Tilknyttet:
    Forfatter
    ved Institutt for sirkulasjon og bildediagnostikk ved Norges teknisk-naturvitenskapelige universitet
1 - 5 av 5